Series 2280 Precision DC Power Supply Reference Manual  Section 7: SCPI command reference 
077085503 / March 2019  7-51 
 
INITiate subsystem 
The commands in the INITiate subsystem control the trigger model. 
:INITiate[n]:CONTinuous 
This command enables and disables continuous triggering. 
Recall settings 
Instrument reset 
Power cycle 
Usage 
:INITiate[n]:CONTinuous <state> 
:INITiate[n]:CONTinuous? 
 
Channel number: For single-channel instruments, the value is 1 
Disable continuous triggering: OFF or 0 
Enable continuous triggering: ON or 1 
Details 
With continuous initiation enabled, sending :INITiate causes error -213, "init ignored." 
Example 
 
ARM:SOUR IMM   
TRIG:SOUR IMM   
INIT:CONT OFF    
ABORt       
ARM:COUN 1   
TRIG:COUN 1   
TRIG:SAMP:COUN 100   
:OUTP ON 
INIT        
Select IMMediate arm source. 
Select IMMediate trigger source. 
Disable continuous initiation. 
Abort operation. Places the instrument in idle. 
Set arm count to 1. 
Set trigger count to 1. 
Set Measure count to 100. 
Turn on the output. 
Start measurement process. 
Also see 
Trigger model (on page 3-25)