Section 2: General operation Series 2280 Precision DC Power Supply Reference Manual
2-44 077085503 / March 2019
Figure 29: Reversed sense leads error
To clear the error, you need to clear the Questionable Instrument Summary Event Register bit and
initiate the trigger model using the front panel or SCPI commands. For more information on registers,
refer to Questionable Instrument Summary Event Register (on page D-16).
To clear an overprotection error using the front panel:
When the overprotection error dialog box is displayed, you can press the ENTER key to initiate the
trigger model.
To clear an overprotection error using a SCPI command:
When an overprotection error occurs, send the following command to clear the register and initiate
the trigger model:
:OUTPut:PROTection:CLEar
Remote communications interfaces
You can choose from one of several communication interfaces to send commands to and receive
responses from the Series 2280.
The Series 2280 can be controlled from only one communication interface at a time. The first
interface from which it receives a message takes control of the instrument. To remove control from
the present interface, send the following command:
:SYSTem:COMMunication:ABORt
The Series 2280 powers up using the last remote interface used.

Table of Contents

Other manuals for Keithley 2280S-32-6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2280S-32-6 and is the answer not in the manual?

Keithley 2280S-32-6 Specifications

General IconGeneral
BrandKeithley
Model2280S-32-6
CategoryPower Supply
LanguageEnglish

Related product manuals