EasyManuals Logo

Keithley 2280S-32-6 Reference Manual

Keithley 2280S-32-6
457 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #388 background imageLoading...
Page #388 background image
Appendix B: Verification Series 2280 Precision DC Power Supply Reference Manual
B-22 077085503 / March 2019
Figure 104: Electronic load and voltmeter connections
Check current noise (20 MHz):
1. Set the voltmeter as follows:
ï‚§ Set to measure DC voltage.
ï‚§ Set the current range to autorange.
2. Set the electronic load as follows:
ï‚§ Set to constant current (CC) mode.
ï‚§ Set to draw a constant current at the test current specified for the DUT in the DC voltage setting
accuracy without remote sense (on page B-4) table.
3. Set the DUT to the full-scale output current.
4. Set the DUT to 0% of the full-scale output voltage (0 V).
5. Turn the DUT output on.
6. Enter the voltmeter reading in the DC voltage setting accuracy without remote sense (on page B-
4) table.
7. Increase the DUT output voltage by 25% of the full-scale output voltage.
8. Repeat steps 5 through 8 until you complete testing at 100% of full-scale output voltage.