Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-100 077085503 / March 2019
:STATus:MEASurement:INSTrument:ISUMmary:CONDition?
This command reads the Measurement Instrument Summary Condition Register of the status model.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:STATus:MEASurement:INSTrument:ISUMmary:CONDition?
Details
This command reads the contents of the Measurement Instrument Summary Condition Register.
For details about interpreting the value of a register, see Measurement Instrument Summary Event
Register (on page D-19).
Example
:STAT:MEAS:INST:ISUM:COND?
Returns the contents of the Measurement
Instrument Summary Condition Register of
the status model.
Also see
None
:STATus:OPERation[:EVENt]?
This command reads and clears the Operation Event Register of the status model.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:STATus:OPERation[:EVENt]?
Details
This command reads the Operation Event Register of the status model and then clears the register.
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in
the register. For details, see Operation Event Register (on page D-11)
Also see
None