EasyManuals Logo

Keithley 2280S-32-6 Reference Manual

Keithley 2280S-32-6
457 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #301 background imageLoading...
Page #301 background image
Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-100 077085503 / March 2019
:STATus:MEASurement:INSTrument:ISUMmary:CONDition?
This command reads the Measurement Instrument Summary Condition Register of the status model.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:STATus:MEASurement:INSTrument:ISUMmary:CONDition?
Details
This command reads the contents of the Measurement Instrument Summary Condition Register.
For details about interpreting the value of a register, see Measurement Instrument Summary Event
Register (on page D-19).
Example
:STAT:MEAS:INST:ISUM:COND?
Returns the contents of the Measurement
Instrument Summary Condition Register of
the status model.
Also see
None
:STATus:OPERation[:EVENt]?
This command reads and clears the Operation Event Register of the status model.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:STATus:OPERation[:EVENt]?
Details
This command reads the Operation Event Register of the status model and then clears the register.
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in
the register. For details, see Operation Event Register (on page D-11)
Also see
None

Table of Contents

Other manuals for Keithley 2280S-32-6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2280S-32-6 and is the answer not in the manual?

Keithley 2280S-32-6 Specifications

General IconGeneral
BrandKeithley
Model2280S-32-6
CategoryPower Supply
LanguageEnglish

Related product manuals