Series 2280 Precision DC Power Supply Reference Manual Section 7: SCPI command reference
077085503 / March 2019 7-149
TRIGger subsystem
The commands in this subsystem configure and control the trigger model for trigger layer.
:TRIGger[:SEQuence[n]]:COUNt
This command sets the number of times to repeat the trigger event.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
1
Usage
:TRIGger[:SEQuence[n]]:COUNt <count>
:TRIGger[:SEQuence[n]]:COUNt?
:TRIGger[:SEQuence[n]]:COUNt? DEFault
:TRIGger[:SEQuence[n]]:COUNt? MINimum
:TRIGger[:SEQuence[n]]:COUNt? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<count>
The number of times to repeat the trigger: 1 to 2500
Example
:TRIG:COUN 4
Set the trigger count to four.
Also see
:ARM[:SEQuence[n]]:COUNt (on page 7-19)
Trigger model (on page 3-25)