:CALibration:PROTected:CODE
This command changes the calibration password.
Usage
:CALibration:PROTected:CODE <string>
Calibration password: An 8-character string including letters and numbers.
Details
Before changing the password, you must first enable calibration mode by using the following
command:
:CAL:PROT:STAT ON,<password>
Example
:CAL:PROT:CODE '12345678'
Change the calibration password to "12345678".
Also see
:CALibration:PROTected:STATe (on page 7-44)
:CALibration:PROTected:COUNt?
This query command returns the number of times that the instrument has completed calibration.
Usage
:CALibration:PROTected:COUNt?
Details
Returns the number of times that the instrument has completed calibration. The calibration count will
increase after saving the calibration constants.
Example
Return the number of times the calibration was
completed.
Also see
None