EasyManuals Logo

Keithley 2280S-32-6 Reference Manual

Keithley 2280S-32-6
457 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #298 background imageLoading...
Page #298 background image
Series 2280 Precision DC Power Supply Reference Manual Section 7: SCPI command reference
077085503 / March 2019 7-97
:STATus:MEASurement:INSTrument:ENABle
This command sets or reads the contents of the Measurement Instrument Event Enable Register of the status
model.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0
Usage
:STATus:MEASurement:INSTrument:ENABle <n>
:STATus:MEASurement:INSTrument:ENABle?
[n]
The status of the Measurement Instrument Enable Register.
Details
This command sets or reads the contents of the enable register of the Measurement Instrument
Event Register.
When one of these bits is set and the corresponding bit in the Measurement Instrument Event
Register is set, the ISUM bit in the Measurement Event Register is set.
When sending binary values, preface <n> with #b. When sending hexadecimal values, preface <n>
with #h. No preface is needed when sending decimal values.
Example
:STAT:MEAS:INST:ENAB #b0000000000000010
Sets the bit 1 of the Measurement
Instrument Enable Register using a
binary value.
You could also send the decimal value 2
or the hexadecimal value #h2.
Also see
Measurement Instrument Event Register (on page D-18)

Table of Contents

Other manuals for Keithley 2280S-32-6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2280S-32-6 and is the answer not in the manual?

Keithley 2280S-32-6 Specifications

General IconGeneral
BrandKeithley
Model2280S-32-6
CategoryPower Supply
LanguageEnglish

Related product manuals