Series 2280 Precision DC Power Supply Reference Manual Appendix C: Calibration and adjustment
077085503 / March 2019 C-7
Voltage calibration connections
For voltage calibration, connect the digital multimeter (DMM) to the Series 2280. For the DMM
specifications, see Required equipment (on page C-3).
Figure 115: Voltage calibration connections
When you are using a binding post on the front panel to make a calibration connection, insert the
jumper wires to short the output terminals and sense terminals on the rear panel. For details, see the
following figure.
Figure 116: Install the output mating connector for 4-wire operation