Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-110 077085503 / March 2019
Example
:STAT:QUES:INST:ISUM:ENAB #b0000000000001010
Sets the overvoltage protection (OVP)
and constant voltage (CV) bits of the
Questionable Instrument Summary
Enable Register using a binary value.
You could also send the decimal value
10 or the hexadecimal value #hA.
Also see
Questionable Instrument summary Event Register (on page D-16)
:STATus:QUEStionable:INSTrument:ISUMmary:CONDition?
This command reads the Questionable Instrument Summary Condition Register of the status model.
Usage
:STATus:QUEStionable:INSTrument:ISUMmary:CONDition?
Details
This command reads the contents of the Questionable Instrument Summary Condition Register.
For details about interpreting the value of a register, see Questionable Instrument Summary Event
Register (on page D-16).
Example
:STAT:QUES:INST:ISUM:COND?
Reads the Questionable Instrument
Summary Condition Register.
Also see
None