Series 2280 Precision DC Power Supply Reference Manual  Section 7: SCPI command reference 
077085503 / March 2019  7-73 
 
:SENSe[n]:<function>:RESolution 
This command sets the measurement resolution for the selected measurement function. 
Recall settings 
Instrument reset 
Power cycle 
Usage 
:SENSe[n]:<function>:RESolution <value> 
 
Channel number: For single-channel instruments, the value is 1 
The measurement function to which this setting applies: 
  Concurrent: CONCurrent[:DC] 
  Current: CURRent[:DC] 
  Voltage: VOLTage[:DC] 
4.5 digit resolution: 4 
5.5 digit resolution: 5 
6.5 digit resolution: 6 
Details 
This command affects how the reading for a measurement is displayed on the front panel of the 
instrument. It also affects the accuracy or speed of measurements. 
When a resolution is selected, the number of power line cycles (NPLCs), autozero, digit, source 
delay, and filter are automatically set to the value described in the following table. 
 
4.5 digits display 
resolution 
  Filter count: 10  
  Filter window: 0.01% 
  Filter state: On 
5.5 digits display 
resolution 
  Filter count: 10  
  Filter window: 0.01% 
  Filter state: Off 
6.5 digits display 
resolution 
  Filter count: 10  
  Filter window: 0.01% 
  Filter state: Off 
Example 
 
Set the 5½ digit measurement resolution  
Also see 
:SENSe[n]:<function>:DIGits (on page 7-65)