Series 2280 Precision DC Power Supply Reference Manual Section 7: SCPI command reference
077085503 / March 2019 7-57
:OUTPut:PROTection:CLEar
This command clears the overvoltage, overcurrent, overtemperature, or reverse fault status.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:OUTPut:PROTection:CLEar
:OUTPut:PROTection:CLEar <channel>
<channel>
The channel number (default value is CH1):
ï‚§ Channel 1: CH1
Channel 2: CH2
Details
Clears the protection status if an overvoltage, overcurrent, or overtemperature protection condition
occurred or a reverse sense fault was detected. For all four faults, this command also clears the
related Questionable Instrument Summary Condition Register bit.
Example
:OUTP:PROT:CLE
Clear the protection status.
Also see
:OUTPut:PROTection:TRIPped? (on page 7-58)
Status model (on page D-1)