Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-44 077085503 / March 2019
:CALibration:PROTected:STATe
This command enables or disables calibration procedures when performing these procedures over the remote
interface.
Instrument reset
Power cycle
Usage
:CALibration:PROTected:STATe <state>, <password>
:CALibration:PROTected:STATe?
Enable the calibration mode: ON or 1
Disable the calibration mode: OFF or 0
The password for calibration:
ï‚§ The maximum length is 8 characters
ï‚§ Not case sensitive
Details
Use this command to enable the calibration mode. After enabled, you can configure and control
calibration operations. If the calibration mode is disabled, the commands related to calibration are
ignored.
When using this command, you are required to provide the calibration password. You could set a new
password by the command:
:CALibration:PROTected:CODE
During the calibration process, changing the calibration mode from 1 (ON) to 0 (OFF) could cause the
instrument to lose the calibration data if you have not saved them with
the :CALibration:PROTected:SAVE command.
Example
:CAL:PROT:STAT ON, 'KI002280'
Enable the calibration procedures.
Also see
None