Section 7: SCPI command reference  Series 2280 Precision DC Power Supply Reference Manual 
7-96  077085503 / March 2019 
 
Example 
 
:STAT:MEAS:ENAB #b0010000000000000 
Sets the ISUM bit of the measurement 
status enable register using a binary 
value. 
You could also send the decimal value 
8192 or the hexadecimal value: #h2000. 
Also see 
Measurement Event Register (on page D-17) 
 
:STATus:MEASurement:INSTrument[:EVENt]? 
This command reads and clears the Measurement Instrument Event Register of the status model. 
Usage 
:STATus:MEASurement:INSTrument[:EVENt]? 
Details 
This attribute reads the Measurement Instrument Event Register of the status model and then clears 
the register. 
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in 
the register. For details, see Measurement Instrument Event Register (on page D-18). 
Also see 
None