Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-62 077085503 / March 2019
:SENSe[n]:<function>:AVERage[:STATe]
This command enables or disables the averaging filter for measurements of the selected function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
1 (ON)
Usage
:SENSe[n]:<function>:AVERage[:STATe] <state>
:SENSe[n]:<function>:AVERage[:STATe]?
[n]
Channel number: For single-channel instruments, the value is 1
<function>
The measurement function to which this setting applies:
ï‚§ Concurrent: CONCurrent[:DC]
ï‚§ Current: CURRent[:DC]
ï‚§ Voltage: VOLTage[:DC]
<state>
The filter status; set to one of the following values:
ï‚§ Disable the averaging filter: OFF or 0
ï‚§ Enable the averaging filter: ON or 1
Details
This command enables or disables the averaging filter. When this is enabled, the reading returned by
the instrument is an averaged value, taken from multiple measurements. The settings of the filter
count and filter type for the selected measure function determines how the reading is averaged.
Example
SENS:CONC:AVER:WIND 0.1
Set the average filter window as 0.1%.
Also see
:SENSe[n]:<function>:AVERage:COUNt (on page 7-61)
:SENSe[n]:<function>:AVERage:TCONtrol (on page 7-63)
:SENSe[n]:<function>:AVERage:WINDow (on page 7-64)

Table of Contents

Other manuals for Keithley 2280S-32-6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2280S-32-6 and is the answer not in the manual?

Keithley 2280S-32-6 Specifications

General IconGeneral
BrandKeithley
Model2280S-32-6
CategoryPower Supply
LanguageEnglish

Related product manuals