EasyManuals Logo

Keithley 2280S-32-6 Reference Manual

Keithley 2280S-32-6
457 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #347 background imageLoading...
Page #347 background image
Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-146 077085503 / March 2019
:TRACe[n]:TRIGger:OCCur?
This command queries whether a level trigger has occurred.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:TRACe[n]:TRIGger:OCCur?
Example
TRAC:TRIG:OCC?
Queries whether a level trigger has occurred.
If a level trigger has occurred, the value of 1 will be
returned.
Also see
None
:TRACe[n]:TRIGger:OFFSet
This command defines the number of points at which the instrument begins to log data in the buffer.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0
Usage
:TRACe[n]:TRIGger:OFFSet <value>
:TRACe[n]:TRIGger:OFFSet?
<value>
From 0 to 100
Example
TRAC:TRIG:OFFS 10
TRAC:TRIG:OFFS?
10 points before the level trigger activation are not cleared.
Also see
None

Table of Contents