Series 2280 Precision DC Power Supply Reference Manual  Section 7: SCPI command reference 
077085503 / March 2019  7-7 
 
*RST 
This command resets the instrument settings to their default values. 
Usage 
*RST 
Details 
Returns the instrument to default settings, cancels all pending commands, and cancels the response 
to any previously received *OPC and *OPC? commands. 
Also see 
Reset default values (on page 2-88) 
 
*SAV 
This command saves the present instrument settings as a user-saved setup. 
Usage 
*SAV <n> 
 
Details 
Save the present instrument settings as a user-saved setup. You can restore the settings with the 
*RCL command. 
Any command that is affected by the*RST command can be saved with the *SAV command. 
Any settings that were stored previously as <n> are overwritten. 
If you view scripts from the front panel of the instrument, these are stored as SETUP<n>. 
Example 
 
Saves the instrument settings in memory location 2. 
Also see 
*RCL (on page 7-6)