Series 2280 Precision DC Power Supply Reference Manual Section 7: SCPI command reference
077085503 / March 2019 7-111
SYSTem subsystem
This subsystem contains commands that affect the overall operation of the instrument, such as
autozero, beeper, communications, event logs, and time.
:SYSTem:AZERo[n][:STATe]
This command enables and disables the autozero.
Recall settings
Instrument reset
Power cycle
Usage
:SYSTem:AZERo[n][:STATe] <state>
:SYSTem:AZERo[n][:STATe]?
Channel number: For single-channel instruments, the value is 1
Disable autozero: OFF or 0
Enable autozero: ON or 1
Details
When autozero is disabled, the reference and internal temperature measurements are not performed.
This increases the measurement speed. However, the reference and temperature reference points
will eventually drift, resulting in inaccurate readings of the input signal. For details, refer to Autozero
measurements (on page 4-5).
Example
Disables autozero to increase measurement speed.
Also see
None