Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-142 077085503 / March 2019
:TRACe[n]:POINts
This command contains the number of readings a buffer can store.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
100
Usage
:TRACe[n]:POINts <newSize>
:TRACe[n]:POINts?
:TRACe[n]:POINts? DEFault
:TRACe[n]:POINts? MINimum
:TRACe[n]:POINts? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<newSize>
The new size for the buffer: 2 to 2500
Details
Use this command to read the number of measurements that a buffer can store.
Example
:TRAC:POIN?
Query the capacity of the buffer.
Output:
1000
Also see
None
:TRACe[n]:POINts:ACTual?
This command queries the number of readings presently stored in the buffer.
Type
Affected by
Where saved
Default value
Query only
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
:TRACe[n]:POINts:ACTual?
[n]
Channel number: For single-channel instruments, the value is 1
Example
:TRAC:POIN:ACT?
Query the number of stored reading in the buffer.
Output:
200
Also see
None