Section 7: TSP command reference Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-78 2657A-901-01 Rev. B/December 2012
Usage
func = display.smuX.measure.func
display.smuX.measure.func = func
func
One of the following values:
• 0 or display.MEASURE_DCAMPS: Selects current measurement function
• 1 or display.MEASURE_DCVOLTS: Selects volts measurement function
• 2 or display.MEASURE_OHMS: Selects ohms measurement function
• 3 or display.MEASURE_WATTS: Selects power measurement function
X
Source-measure unit (SMU) channel (for example,
display.smua.measure.func applies to SMU channel A)
Details
Selects the displayed measurement function: Amperes, volts, ohms, or watts.
Example
display.smua.measure.func = display.MEASURE_DCAMPS
Selects the current measure function
for SMU A.
Also see
display.smuX.limit.func (on page 7-77)
display.trigger.clear()
This function clears the front-panel trigger event detector.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
display.trigger.clear()
Details
The trigger event detector remembers if an event has been detected since the last display.trigger.wait()
call. This function clears the trigger event detector and discards the previous history of TRIG key presses.
This attribute also clears the display.trigger.overrun attribute.
Also see
display.trigger.overrun (on page 7-79)
display.trigger.wait()
(on page 7-80)