EasyManuals Logo

Keithley 2657A User Manual

Keithley 2657A
805 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #524 background imageLoading...
Page #524 background image
Model 2657A High Power System SourceMeter® Instrument Reference Manual Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012 7-211
smuX.reset()
This function turns off the output and resets the SMU to the default settings.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
smuX.reset()
X
Source-measure unit (SMU) channel (for example, smua.reset() applies to SMU
channel A)
Details
This function turns off the output and returns the specified SMU to its default settings.
Example
smua.reset()
Turns off the output and resets SMU
channel A to its default settings.
Also see
reset() (on page 7-155)
smuX.savebuffer()
This function saves one source-measure unit (SMU) dedicated reading buffer to nonvolatile memory (there are
two dedicated reading buffers for each SMU).
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
smuX.savebuffer(smuX.nvbufferY)
X
SMU channel (for example, smua.savebuffer(smua.nvbuffer1) applies to
SMU channel A)
Y
SMU dedicated reading buffer (1 or 2)
Details
When the instrument is turned off and back on, the dedicated reading buffers are restored from nonvolatile
memory to their last saved values.
Example
smua.savebuffer(smua.nvbuffer1)
Saves buffer 1 (SMU channel A) to
internal memory.
Also see
Reading buffers (on page 3-6)
savebuffer()
(on page 7-156)
smuX.nvbufferY
(on page 7-210)