Section 7: TSP command reference Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-86 2657A-901-01 Rev. B/December 2012
Example
print(eventlog.enable)
eventlog.enable = eventlog.DISABLE
print(eventlog.enable)
Displays the present status of the Model
2657A event log.
Output:
1.00000e+00
0.00000e+00
Also see
eventlog.all() (on page 7-84)
eventlog.clear()
(on page 7-84)
eventlog.count
(on page 7-85)
eventlog.next()
(on page 7-86)
eventlog.overwritemethod
(on page 7-87)
eventlog.next()
This function returns the oldest message from the event log and removes it from the log.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
logString = eventlog.next()
logString
The next log entry
Details
Returns the next entry from the event log and removes it from the log.
If there are no entries in the event log, returns the value nil.
Example 1
print(eventlog.next())
Get the oldest message in the event log and remove that entry from the log.
Output:
17:28:22.085 10 Oct 2009, LAN2, 192.168.1.102, LXI, 0, 1192037134, <no time>, 0,
0x0
Example 2
print(eventlog.next())
If you send this command when there is nothing in the event log, you will get the following output:
nil
Also see
eventlog.all() (on page 7-84)
eventlog.clear()
(on page 7-84)
eventlog.count
(on page 7-85)
eventlog.enable
(on page 7-85)
eventlog.overwritemethod
(on page 7-87)

Table of Contents