Model 2657A High Power System SourceMeter® Instrument Reference Manual Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012 7-341
trigger.timer[N].reset()
This function resets some of the trigger timer settings to their factory defaults.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
trigger.timer[N].reset()
N
Trigger timer number (1 to 8)
Details
The trigger.timer[N].reset() function resets the following attributes to their factory defaults:
trigger.timer[N].count
trigger.timer[N].delay
trigger.timer[N].delaylist
trigger.timer[N].passthrough
trigger.timer[N].stimulus
It also clears trigger.timer[N].overrun.
Example
trigger.timer[1].reset()
Resets the attributes associated with timer 1 back
to factory default values.
Also see
trigger.timer[N].count (on page 7-337)
trigger.timer[N].delay
(on page 7-338)
trigger.timer[N].delaylist
(on page 7-338)
trigger.timer[N].overrun
(on page 7-339)
trigger.timer[N].passthrough
(on page 7-340)
trigger.timer[N].stimulus
(on page 7-341)
trigger.timer[N].stimulus
This attribute specifies which event starts the timer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
Recall setup
Trigger timer N reset
Not saved 0
Usage
eventID = trigger.timer[N].stimulus
trigger.timer[N].stimulus = eventID
eventID
The event that triggers the timer delay
N
Trigger timer number (1 to 8)