Model 2657A High Power System SourceMeter® Instrument Reference Manual Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012 7-349
tsplink.trigger[N].clear()
This function clears the event detector for a trigger.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
tsplink.trigger[N].clear()
N
The trigger line (1 to 3)
Details
The trigger event detector enters the detected state when an event is detected.
tsplink.trigger[N].clear() clears a trigger event detector, discards the history of the trigger line, and
clears the tsplink.trigger[N].overrun attribute.
Example
tsplink.trigger[2].clear()
Clears trigger event on synchronization line 2.
Also see
tsplink.trigger[N].mode (on page 7-350)
tsplink.trigger[N].overrun
(on page 7-352)
tsplink.trigger[N].release()
(on page 7-353)
tsplink.trigger[N].stimulus
(on page 7-354)
tsplink.trigger[N].wait()
(on page 7-356)
tsplink.trigger[N].EVENT_ID
This constant identifies the number that is used for the trigger events.
Type TSP-Link accessible
Affected by Where saved Default value
Constant Yes
Usage
eventID = tsplink.trigger[N].EVENT_ID
eventID
The trigger event number
N
The trigger line (1 to 3)