Model 2657A High Power System SourceMeter® Instrument Reference Manual Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012 7-175
setup.recall()
This function recalls settings from a saved setup.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
setup.recall(id)
id
An integer or string that specifies the location of the setup to recall:
Factory default setup (0)
User-saved setup in nonvolatile memory (1 to 5 )
User-saved setup on a USB flash drive ("/path/filename")
Details
When the id parameter is an integer (n), it is interpreted as the setup number to restore from the instrument's
nonvolatile memory. When n = 0, the instrument recalls the factory default setup; when n = 1 to 5, the
instrument recalls a user-saved setup.
When the id parameter is a string, it is interpreted as the path and file name of the setup to restore from a file on
a USB flash drive. The path may be absolute or relative to the current working directory.
Before a setup is recalled, an instrument reset is performed.
Example 1
setup.recall(1)
Recall the user-saved setup at location 1.
Example 2
setup.recall("/usb1/KEITHLEY_30730.set")
Recall a user-saved setup stored in a file
named KEITHLEY_30730 on a USB flash
drive.
Also see
Saved setups (on page 2-38)
setup.save()
(on page 7-175)
setup.save()
This function saves the present setup as a user-saved setup.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
setup.save(id)
id
An integer or string specifying where to save the user setup:
Save in nonvolatile memory (1 to 5)
Save as user-saved setup on a USB flash drive ("/path/filename")