Model 2657A High Power System SourceMeter® Instrument Reference Manual Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012 7-333
trigger.blender[N].reset()
This function resets some of the trigger blender settings to their factory defaults.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
trigger.blender[N].reset()
N
The trigger event blender (1 to 4)
Details
The trigger.blender[N].reset() function resets the following attributes to their factory defaults:
• trigger.blender[N].orenable
• trigger.blender[N].stimulus[M]
It also clears trigger.blender[N].overrun.
Example
trigger.blender[1].reset()
Resets the trigger blender 1
settings to factory defaults.
Also see
trigger.blender[N].orenable (on page 7-331)
trigger.blender[N].overrun
(on page 7-332)
trigger.blender[N].stimulus[M]
(on page 7-333)
trigger.blender[N].stimulus[M]
This attribute specifies which events trigger the blender.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
Recall setup
Trigger blender N reset
Not saved 0
Usage
eventID = trigger.blender[N].stimulus[M]
trigger.blender[N].stimulus[M] = eventID
eventID
The event that triggers the blender action; see Details
N
An integer representing the trigger event blender (1 to 4)
M
An integer representing the stimulus index (1 to 4)
Details
There are four stimulus inputs that can each select a different event. The eventID parameter can be the event
ID of any trigger event.
Use zero to disable the blender input.