Model 2657A High Power System SourceMeter® Instrument Reference Manual Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012 7-77
Also see
Display resolution (on page 3-69)
display.smuX.limit.func
This attribute specifies the type of limit value setting displayed.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
Recall setup
Saved setup 0 (display.LIMIT_IV)
Usage
func = display.smuX.limit.func
display.smuX.limit.func = func
func
One of the following values:
0 or display.LIMIT_IV: Displays the primary limit setting
1 or display.LIMIT_P: Displays the power limit setting
X
Source-measure unit (SMU) channel (for example, display.smua.limit.func
applies to SMU channel A)
Details
Selects the displayed limit function: primary (IV) or power (P).
Example
display.smua.limit.func = display.LIMIT_P
Specifies the power limit value is
displayed for SMU Channel A.
Also see
display.smuX.measure.func (on page 7-77)
display.smuX.measure.func
This attribute specifies the type of measurement that is being displayed.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
Recall setup
Saved setup 1 (display.MEASURE_DCVOLTS)