Model 2657A High Power System SourceMeter® Instrument Reference Manual Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012 7-345
tsplink.master
This attribute reads the node number assigned to the master node.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes Not applicable Not applicable Not applicable
Usage
masterNodeNumber = tsplink.master
masterNodeNumber
The node number of the master node
Details
After doing a TSP-Link reset (tsplink.reset()), use this attribute to access the node number of the master in
a set of instruments connected over TSP-Link.
Example
LinkMaster = tsplink.master
Store the TSP-Link master
node number in a variable
called LinkMaster.
Also see
tsplink.reset() (on page 7-347)
tsplink.node
This attribute defines the node number.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Not applicable Nonvolatile
memory
1
Usage
nodeNumber = tsplink.node
tsplink.node = nodeNumber
nodeNumber
Set node to a number (1 to 64)
Details
This attribute sets the TSP-Link node number and saves the value in nonvolatile memory.
Changes to the node number do not take effect until the next time tsplink.reset() is executed on any node
in the system.
Each node connected to the TSP-Link system must be assigned a different node number.
Example
tsplink.node = 2
Sets the TSP-Link node to number 2.
Also see
tsplink.reset() (on page 7-347)
tsplink.state
(on page 7-347)