Model 2657A High Power System SourceMeter® Instrument Reference Manual  Section 7: TSP command reference
 
2657A-901-01 Rev. B/December 2012  7-365 
 
Details 
This function sets and gets the termination character sequence that is used to indicate the end of a line for a 
TSP-Net connection. 
Using the termSequence parameter sets the termination sequence. The present termination sequence is 
always returned. 
For the termSequence parameter, use the same values listed in the table above for type. There are four 
possible combinations, all of which are made up of line feeds (LF or 0x10) and carriage returns (CR or 0x13). For 
TSP-enabled devices, the default is tspnet.TERM_LF. For devices that are not TSP-enabled, the default is 
tspnet.TERM_CRLF. 
 
Example 
 
deviceID = tspnet.connect("192.0.2.1") 
if deviceID then 
   tspnet.termination(deviceID, 
tspnet.TERM_LF) 
end 
Sets termination type for IP address 
192.0.2.1 to TERM_LF. 
Also see 
tspnet.connect() (on page 7-359) 
tspnet.disconnect()
 (on page 7-360) 
 
tspnet.timeout 
This attribute sets the timeout value for the tspnet.connect(), tspnet.execute(), and tspnet.read() 
commands. 
Type  TSP-Link accessible  Affected by  Where saved  Default value 
Attribute (RW)  Yes  Instrument reset
Recall setup 
Not saved      20.0 (20 s) 
 
Usage 
value = tspnet.timeout 
tspnet.timeout = value 
 
value 
The timeout duration in seconds (0.001 s to 30.000 s) 
 
Details 
This attribute sets the amount of time the tspnet.connect(), tspnet.execute(), and tspnet.read() 
commands will wait for a response. 
The time is specified in seconds. The timeout may be specified to millisecond resolution, but is only accurate to 
the nearest 10 ms. 
 
Example 
 
tspnet.timeout = 2.0 
Sets the timeout duration to two seconds. 
 
Also see 
tspnet.connect() (on page 7-359) 
tspnet.execute()
 (on page 7-361) 
tspnet.read()
 (on page 7-362)