Model 2657A High Power System SourceMeter® Instrument Reference Manual Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012 7-91
Details
This function may buffer data until a flush (fileVar:flush() or io.flush()) or close (fileVar:close()
or io.close()) operation is performed.
Also see
fileVar:close() (on page 7-88)
fileVar:flush()
(on page 7-88)
io.close()
(on page 7-103)
io.flush()
(on page 7-104)
io.open()
(on page 7-105)
format.asciiprecision
This attribute sets the precision (number of digits) for all numbers printed with the ASCII format.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) No Instrument reset
Recall setup
Not saved 6
Usage
precision = format.asciiprecision
format.asciiprecision = precision
precision
A number representing the number of digits to be printed for numbers printed with
the print(), printbuffer(), and printnumber() functions; must be a
number between 1 and 16
Details
This attribute specifies the precision (number of digits) for numeric data printed with the print(),
printbuffer(), and printnumber() functions. The format.asciiprecision attribute is only used with
the ASCII format. The precision value must be a number between 1 and 16.
Note that the precision is the number of significant digits printed. There is always one digit to the left of the
decimal point; be sure to include this digit when setting the precision.
Example
format.asciiprecision = 10
x = 2.54
printnumber(x)
format.asciiprecision = 3
printnumber(x)
Output:
2.540000000e+00
2.54e+00
Also see
format.byteorder (on page 7-92)