Model 2657A High Power System SourceMeter® Instrument Reference Manual Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012 7-139
Details
When set to 1, the IEEE Std 488.2 common commands generate prompts if prompting is enabled with the
localnode.prompts attribute. If set to 1, limit the number of *trg commands sent to a running script to 50
regardless of the setting of the localnode.prompts attribute.
When set to 0, IEEE Std 488.2 common commands will not generate prompts. When using the *trg command
with a script that executes trigger.wait() repeatedly, set localnode.prompts4882 to 0 to avoid
problems associated with the command interface input queue filling.
This attribute resets to the default value each time the instrument power is cycled.
When using this command from a remote node, localnode should be replaced with the node reference, for
example node[5].prompts4882.
Example
localnode.prompts4882 = 0
Disables IEEE Std 488.2 common command prompting.
Also see
localnode.prompts (on page 7-137)
localnode.reset()
This function resets the local node instrument.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
localnode.reset()
Details
If you want to reset a specific instrument or a subordinate node, use the node[X].reset() command.
A local node reset includes:
• Source-measure unit (SMU) attributes affected by a SMU reset are reset
• Other settings are restored back to factory default settings
A localnode.reset() is different than a reset() because reset() resets the entire system.
When using this command from a remote node, localnode should be replaced with the node reference, for
example node[5].reset().
Example
localnode.reset()
Resets the local node.
Also see
reset() (on page 7-155)
smuX.reset()
(on page 7-211)
localnode.revision
This attribute stores the firmware revision level.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes Not applicable Not applicable Not applicable

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2657A and is the answer not in the manual?

Keithley 2657A Specifications

General IconGeneral
BrandKeithley
Model2657A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals