Section 7: Triggering Model 6517B Electrometer Reference Manual
7-18 6517B-901-01 Rev. C / August 2015
For typical asynchronous trigger link operation, the measure layer is configured with measure source
set to TRIGLINK and Trigger link mode set to ASYNCHRONOUS. You must also select input and
output lines for the measure layer. Input and output triggers can be set to any of the six lines, but they
cannot use the same line. For example, if you select line #1 for input triggers, then output triggers
must use one of the other five lines (#2 through #6).
During operation in the measure layer, each trigger link input trigger makes a measurement. After the
user-programmed DELAY and the measurement settling time, the Model 6517B outputs a trigger link
completion pulse (typically to a scanner to close the next channel). The measure layer is configured
using the CONFIGURE TRIGGER menu. See Trigger configuration menu (on page 7-2) for more
information.
The scan layer and arm layer can also be programmed for trigger link, where scan source is set to
TRIGLINK, and Arm source is set to TRIGLINK. When using trigger link in these layers, you must also
select input and output lines as you did in the measure layer. Keep in mind that you can use the same
lines in the scan and arm layers as selected in the measure layer.

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
Input Impedance>200 TΩ
Electrometer TypeElectrometer/High Resistance Meter
Current Measurement Range20 fA to 20 mA
DisplayVacuum Fluorescent Display
Communication InterfacesIEEE-488 (GPIB), RS-232
Power Supply100 V / 120 V / 220 V / 240 V; 50 Hz / 60 Hz
Resistance Measurement RangeUp to 200 TΩ

Related product manuals