Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-156 6517B-901-01 Rev. C / August 2015
General test sequence commands
:ARM
This command arms the selected test sequence.
Usage
:TSEQuence:ARM
Details
This action command is used to arm the selected test sequence. The test starts when the
programmed source event occurs. When using this command for the alternating polarity High
resistance/resistivity transition filters, wait at least 0.3 seconds before using the *TRG trigger
command.
Also see
:TSOurce <name> (on page 14-158)
:TYPE <name> (on page 14-157)
:ABORt
This command stops the test sequence that is in process.
Usage
:TSEQuence:ABORt
Details
This command is used to stop a test sequence that is in process. You can start it again by arming the
test and providing the programmed source event.
Also see
:ARM (on page 14-156)
:TSOurce <name> (on page 14-158)