Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-147
:IMMediate
This command is used to bypass the arm, scan, and measure control source and delays.
Usage
:ARM[:SEQuence[1]][:LAYer[1]]:IMMediate
:ARM[:SEQuence[1]]:LAYer2:IMMediate
:TRIGger:[:SEQuence[1]]:IMMediate
Details
These action commands are used to bypass the specified control source of the trigger model. They
are used when you do not wish to wait for the programmed event. Note from the IEEE-488 trigger
model (on page 11-16) that :ARM:LAY2:IMM and :TRIG:IMM also bypass the delay.
The instrument must be waiting for the appropriate event when this command is sent. Otherwise, an
error occurs and the command is ignored. When the command is sent, the bypass occurs and
operation continues on.
Also see
None
:COUNT <n>
This command is used to set the arm layer count.
Usage
:ARM[:SEQuence[1]][:LAYer[1]]:COUNt <n>
:ARM[:SEQuence[1]]:LAYer2:COUNt <n>
:TRIGger[:SEQuence[1]]:COUNt <n>
:COUNt?
:COUNt? DEFault
:COUNt? MINimum
:COUNt? MAXimum
<n>
1 to 99999: Specify count
INF: Sets count to infinite
DEFault: Sets count to 1
MINimum: Sets count to 1
MAXimum: Sets count to 99999
Details
These commands are used to specify how many times operation loops around in the specified layer
of operation. For example, if the measure count is set to 10, operation continues to loop around in the
measure layer until 10 measurements are performed. After the 10th measurement, operation
proceeds back up to the scan layer.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
Input Impedance>200 TΩ
Electrometer TypeElectrometer/High Resistance Meter
Current Measurement Range20 fA to 20 mA
DisplayVacuum Fluorescent Display
Communication InterfacesIEEE-488 (GPIB), RS-232
Power Supply100 V / 120 V / 220 V / 240 V; 50 Hz / 60 Hz
Resistance Measurement RangeUp to 200 TΩ

Related product manuals