Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-88 6517B-901-01 Rev. C / August 2015
:KSURface <NRf>
This command is used to specify the Ks parameter for surface resistivity.
Usage
[:SENSe[1]]:RESistance:RESistivity:USER:KSURface <NRf>
:KSURface?
<NRf>
0.001 to 999.999: Ks parameter
Details
Description This command is used to specify the Ks parameter for surface resistivity measurements
when using a user-supplied test fixture. This command is not needed when using the Model 8009 test
fixture.
The Ks parameter is calculated as follows:
Ks = P/g
Where: P = Effective perimeter (in mm) of the guarded electrode
g = Distance (mm) between the guarded electrode and the ring electrode
Also see
None
:KVOLume <NRf>
This command is used to specify the Kv parameter for volume resistivity.
Usage
[:SENSe[1]]:RESistance:RESistivity:USER:KVOLume <NRf>
:KVOLume?
<NRf>
0.001 to 999.999: Kv parameter
Details
This command is used to specify the Kv parameter for volume resistivity measurements when using a user-
supplied test fixture. This command is not needed when using the Model 8009 test fixture.
Also see
None
:SOURce subsystem
This subsystem is used to set the logic level (true or false) of each digital output line, and is used to
configure the V-source. Note that the V-source is controlled (standby/operate) from the :OUTput1
subsystem.
The commands for this subsystem are summarized in the :SOURce command summary (on page 14-
20).

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
Input Impedance>200 TΩ
Electrometer TypeElectrometer/High Resistance Meter
Current Measurement Range20 fA to 20 mA
DisplayVacuum Fluorescent Display
Communication InterfacesIEEE-488 (GPIB), RS-232
Power Supply100 V / 120 V / 220 V / 240 V; 50 Hz / 60 Hz
Resistance Measurement RangeUp to 200 TΩ

Related product manuals