Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-165
:MVOLtage <NRf>
This command specifies the measure voltage for the specified test.
Usage
:TSEQuence:SRESistivity:MVOLtage <NRf>
:TSEQuence:VRESistivity:MVOLtage <NRf>
:TSEQuence:SIResistance:MVOLtage <NRf>
:MVOLtage?
<NRf>
-100.0 to 100.0 Volts; 100 V range
-1000 to 1000 Volts; 1000 V range
Details
This command is used to specify the measure voltage for the specified test.
This configuration command is used for the following tests:
Surface resistivity test
Volume resistivity test
Surface insulation resistance test
:TSEQuence:SRESistivity:MVOLtage specifies the measure voltage for the surface resistivity
test.
:TSEQuence:VRESistivity:MVOLtage specifies the measure voltage for the volume resistivity
test.
:TSEQuence:SIResistance:MVOLtage specifies the measure voltage for the surface insulation
resistance test.
MVOLtage? queries the measure voltage.
Also see
None