Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-99
:VSLimit <b>
This command is used to control the 200 V V-source Limit.
Usage
:ROUTe:SCAN:VSLimit <b>
:VSLimit?
<b>
ï‚· 0 or OFF: Disable 200V V-source Limit
ï‚· 1 or ON: Enable 200V V-source Limit
Details
This 200V V-source limit is used to help protect internal scanner cards from high voltage that could
damage them. With the 200V limit enabled, the V-source can never source more than +/-200V.
Note that the :SOURce:VOLTage:LIMit command is also used to set the voltage limit of the V-
source. Of the two voltage limit methods, the one that is set to the lower limit value (absolute) is the
one that is in effect.
Also see
None
:STATus subsystem
The STATus subsystem is used to control the status registers of the Model 6517B. These registers
and the overall status structure is explained in Status structure (on page 13-1). The commands in this
subsystem are summarized in :STATus command summary (on page 14-21).

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals