Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-84 6517B-901-01 Rev. C / August 2015
:LEVel <NRf>
This command is used to specify the discharge level.
Usage
[:SENSe[1]]:CHARge:ADIScharge:LEVel <NRf>
:LEVel?
<NRf>
-2.2e6 to 2.2e6: Specify discharge level
Details
This command is used to specify the coulombs level for auto discharge. Charge resets when the
charge reading reaches the specified level.
Also see
None
:XFEedback <b>
This command is used to control external feedback.
Usage
[:SENSe[1]]:VOLTage[:DC]:XFEedback <b>
:XFEedback?
<b>
ON or 1: Enable external feedback
OFF or 0: Disable external feedback
Details
This command is used to enable or disable external feedback (see Using external feedback (on page 5-10) for
more information).
Also see
None