Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-45
:IMMediate
This command performs CALC 3.
Usage
:CALCulate3:IMMediate
Details
Description When you change the configuration of the limit test, the next reading is evaluated
according to that new test configuration. If the instrument is not in a continuous measurement mode
(for example, waiting for a manual trigger), the test is not performed until the next reading conversion
occurs.
This action command allows you to re-process the current input data to test new limits. For example,
assume the instrument is in a non-continuous measurement mode and requires a manual trigger to
cause the next reading conversion. Changing the test limits does not affect the last test result.
However, sending the :IMMediate command re-processes the data and evaluates the reading
according to the new test limits. Note that sending the :IMMediate command does not initiate a
reading conversion.
Also see
None
DISPlay subsystem
The display subsystem controls the display of the Model 6517B and is summarized in the DISPlay command
summary (on page 14-9).

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals