Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-159
:TLIne <NRf>
These commands specify and query the trigger link line.
Usage
:TSEQuence:TLIne <NRf>
:TLINe?
<NRf>
1: Line #1
2: Line #2
3: Line #3
4: Line #4
5: Line #5
6: Line #6
Details
This command is used to specify the input trigger link line for the TLINk control source. This
command is not used if a different control source is selected.
:TLINe? queries the trigger link line.
Also see
:TSOurce <name> (on page 14-158)
:STARt <NRf>
This command specifies the start voltage for a diode leakage or staircase sweep test.
Usage
:TSEQuence:DLEakage:STARt <NRf>
:TSEQuence:STSWeep:STARt <NRf>
:STARt?
<NRf>
-100.0 to 100.0 Volts; 100V range
-1000 to 1000 Volts; 1000V range
Details
This configuration command is used for the following tests:
Diode leakage test
Staircase sweep test
:TSEQuence:DLEakage:STARt specifies the start voltage for a diode leakage test.
:TSEQuence:STSWeep:STARt specifies the start voltage for a staircase sweep test.
:STARt? queries the start voltage
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
Input Impedance>200 TΩ
Electrometer TypeElectrometer/High Resistance Meter
Current Measurement Range20 fA to 20 mA
DisplayVacuum Fluorescent Display
Communication InterfacesIEEE-488 (GPIB), RS-232
Power Supply100 V / 120 V / 220 V / 240 V; 50 Hz / 60 Hz