Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-127
:DATE <yr>, <mo>, <day>
This command is used to set the date for the clock.
Usage
:SYSTem:DATE <yr>, <mo>, <day>
:DATE?
<yr>
ï‚· 2005 to 2104: Specify year
<mo>
ï‚· 1 to 12: Specify month
<day>
ï‚· 1 to 31: Specify day
Details
This command is used to set the date for the real-time clock. Setting an invalid date (for example,
February 30, 2007) results in an error and the previous date is retained.
Also see
None
:TIME <hr>, <min>, <sec>
This command is used to set the time for the clock.
Usage
:SYSTem:TIME <hr>, <min>, <sec>
:TIME?
<hr>
ï‚· 0 to 23: Specify hour (24 hr format)
<min>
ï‚· 0 to 59: Specify minute
<sec>
ï‚· 0.00 to 59.9: Specify second (rounded to 1/100 seconds)
Details
This command is used to set the time for the real-time clock. Note that the time must be set using the
24-hour format (for example, hour 13 is 1 PM). Setting an invalid time results in an error, and the
previous time is retained.
:TIME? returns the time to the nearest hundredth of a second.
Also see
None
:TSTamp commands
The following commands are used to configure and control the time stamp.

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals