Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-123
:POSetup <name>
This command programs power-on defaults.
Usage
SYSTem:POSetup <name>
SYSTem:POSetup?
<name>
RST: Select *RST defaults on power-up
PRESet: Select :SYSTem:PRESet defaults on power-up
SAVX: Select saved defaults on power-up where X = 0 to 9
Details
This command is used to select or query the power-on defaults. With RST selected, the instrument
powers up to the *RST default conditions. With PRESet selected, the instrument powers up to the
:SYStem:PRESet default conditions. Default conditions are listed in the SCPI tables (refer to the
SCPI command summary (on page 14-6)).
With one of the SAV parameters selected, the instrument powers-on to the setup that is saved in the
specified memory location using the *SAV command.
Also see
:PRESet (on page 14-122)
:VERSion?
This command reads the version of the SCPI standard.
Usage
:SYSTem:VERSion?
Details
This query command is used to read the version of the SCPI standard being used by the Model
6517B.
Example
SYST:VERS
Example output:
1996.0
The above response message indicates the
version of the SCPI standard.
Also see
None