Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-77
:COUNt <n>
This command is used to specify the filter count for volts, amps, ohms, and coulombs.
Usage
[:SENSe[1]]:VOLTage[:DC]:AVERage:COUNt <n>
[:SENSe[1]]:CURRent[:DC]:AVERage:COUNt <n>
[:SENSe[1]:RESistance:AVERage:COUNt <n>
[:SENSe[1]]:CHARge:AVERage:COUNt <n>
:COUNt?
:COUNt? DEFault
:COUNt? MINimum
:COUNt? MAXimum
<n>
1 to 100: specify filter count
DEFault: 10
MINimum: 1
MAXimum: 100
Details
These commands are used to specify the filter count. In general, the filter count is the number of
readings that are acquired and stored in the filter buffer for the averaging calculation. The larger the
filter count, the more filtering that is performed.
Also see
None
[:STATe] <b>
This command is used to control the filter for volts, amps, ohms, and coulombs.
Usage
[:SENSe[1]]:VOLTage[:DC]:AVERage[:STATe] <b>
[:SENSe[1]]:CURRent[:DC]:AVERage[:STATe] <b>
[:SENSe[1]:RESistance:AVERage[:STATe] <b>
[:SENSe[1]]:CHARge:AVERage[:STATe] <b>
[:STATe]?
<b>
0 or OFF: Disable the digital filter
1 or ON: Enable the digital filter
Details
These commands are used to enable or disable the digital averaging filter for the specified function.
When enabled, readings are filtered according to how the filter is configured.
Attempting to set STATe to ON with TYPE at NONE and MED:STAT at OFF generates a "Settings
Conflict Error." Setting STATe to ON during a scan or test sequence changes TCON to REPeat.
Also see
None