EasyManua.ls Logo

Keithley 6517B

Keithley 6517B
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-26 6517B-901-01 Rev. C / August 2015
:TRIGger command summary
Command Description Default parameter SCPI
:INITiate
Subsystem command path:
[:IMMediate]
Initiate one trigger cycle
:CONTinuous <b>
Enable or disable continuous initiation (Note 1)
:CONTinuous?
Query continuous initiation
:POFLag <name>
INCLude or EXCLude no pending flag
for initiate commands
INCLude
:POFLag?
Query pending operation flag
:ABORt
Reset trigger system
:ARM[:SEQuence[1]]
Path to configure arm layers:
[:LAYer[1]]
Path to program Arm Layer 1:
:IMMediate
Loop around control source
:COUNt <n>
Specify arm count (1 to 99999, or INF) 1
:COUNt?
Query arm count
:SOURce <name>
Select control source (HOLD,
IMMediate, RTCLock, MANual, BUS,
TLINk, or EXTernal)
IMMediate
(Note 2)
:SOURce?
Query control source
:SIGNal
Loop around control source
:TCONfigure
Path to configure triggers:
:DIRection
<name>
Enable (SOURce) or disable
(ACCeptor)
ACCeptor
:DIRection?
Query direction
:ASYNchronous
Path to configure asynchronous
trigger link:
:ILINe
<NRf>
Select input line (1 to 6) 2
:ILINe?
Query input line
:OLINe
<NRf>
Select output line (1 to 6) 1
:OLINe?
Query output line
:RTCLock
Path to configure the RTCLock
control source:
:DATE <yr>,
<mo>, <day>
Specify the date for the clock event
:DATE?
Query the date for the clock event
:TIME <hr>,
<min>, <sec>
Specify the time for the clock event
:TIME?
Query the time for the clock event
:LAYer2
Path to program Arm Layer 2:
:IMMediate
Loop around control source
:COUNt <n>
Specify scan count (1 to 99999 or INF) (Note 3)
:COUNt?
Query scan count

Table of Contents

Other manuals for Keithley 6517B

Related product manuals