Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-169
:COUNt <NRf>
This command specifies how many times to repeat the programmed square wave cycle.
Usage
:TSEQuence:SQSWeep:COUNt <NRf>
:COUNt?
<NRf>
1 to MAX/2; see Table 8-2
Details
This command is used to specify the cycle count.
:COUNt? queries cycle count.
Also see
None
:OFSVoltage <NRf>
This command specifies the offset voltage for the alternating polarity resistance/resistivity test.
Usage
:TSEQuence:ALTPolarity:OFSVoltage <NRf>
:OFSVoltage?
<NRf>
-1000 to 1000 Offset voltage
Details
:OFSVoltage? queries the offset voltage.
Also see
None
:ALTVoltage <NRf>
This command specifies the alternating voltage for the alternating polarity resistance/resistivity test.
Usage
:TSEQuence:ALTPolarity:ALTVoltage <NRf>
:ALTVoltage <NRf>
<NRf>
-1000 to 1000 alternating voltage
Details
:ALTVoltage? queries the alternating voltage.
Also see
None