Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-134 6517B-901-01 Rev. C / August 2015
:LLOCkout <b>
This command enables or disables local lockout.
Usage
:SYSTem:LLOCkout <b>
:LLOCkout?
<b>
ï‚· 0 or OFF: Disable local lockout
ï‚· 1 or ON: Enable local lockout
Details
If enabled, the front-panel keys are locked out (non-operational) when the instrument is in remote. If
disabled, the front panel keys are operational in remote. The instrument must be in remote in order to
use this command.
Taking the instrument out of remote restores front-panel key operation, but does not change the
status of the :LLOCkout command.
Note that this command can only be sent over the RS-232 interface.
Also see
:LOCal (on page 14-133)
:REMote (on page 14-133)
Basic trigger commands
The following commands are used to configure the Model 6517B for basic triggering. See Triggering
(on page 7-1) for details on basic triggering.
[:EXECute]
This command exits from advanced triggering and selects basic triggering.
Usage
:SYSTem:MACRo:TRIGger:MODE
Details
When this action command is sent, the instrument exits from advanced triggering and select basic
triggering. Sending this command while already in basic triggering has no effect on the instrument.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals