Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-166 6517B-901-01 Rev. C / August 2015
:MTIMe <NRf>
This command specifies the measure time for the specified test.
Usage
:TSEQuence:SRESistivity:MTIMe <NRf>
:TSEQuence:VRESistivity:MTIMe <NRf>
:TSEQuence:SIResistance:MTIMe <NRf>
:TSEQuence:ALTPolarity:MTIMe <NRf>
:MTIMe?
<NRf>
0 to 9999.9 (seconds)
0.5 to 9999.9 (seconds) for the alternating polarity resistance/resistivity test
Details
This command is used to specify the measure time for the specified test.
This configuration command is used for the following tests:
Surface resistivity test
Volume resistivity test
Surface insulation resistance test
Alternating polarity resistance/resistivity test
:TSEQuence:SRESistivity:MTIMe specifies the measure time for the surface resistivity test.
:TSEQuence:VRESistivity:MTIMe specifies the measure time for the volume resistivity test.
:TSEQuence:SIResistance:MTIMe specifies the measure time for the surface insulation
resistance test.
:TSEQuence:ALTPolarity:MTIMe specifies the measure time for the alternating polarity test.
:MTIMe? queries the measure time of the specified test.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
Input Impedance>200 TΩ
Electrometer TypeElectrometer/High Resistance Meter
Current Measurement Range20 fA to 20 mA
DisplayVacuum Fluorescent Display
Communication InterfacesIEEE-488 (GPIB), RS-232
Power Supply100 V / 120 V / 220 V / 240 V; 50 Hz / 60 Hz
Resistance Measurement RangeUp to 200 TΩ