EasyManuals Logo

Keithley 6517B Reference Manual

Keithley 6517B
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #470 background imageLoading...
Page #470 background image
In this appendix:
Introduction ............................................................................... A-1
Introduction
This section discusses front panel triggering, trigger configuration, and external triggering, including
example setups.
Calculating volts accuracy
From the specifications, volts is calculated as follows:
Accuracy = ± (% rdg + counts)
The following example shows how to compute accuracy for the 2 V range:
Assume that the voltage you are measuring is reading exactly 1.00000 V on the 2 V range.
From the specifications:
Accuracy = ± (0.025 % of 1 V + 4 counts)
= ± (0.00025 V + 4 counts)
= ± (0.00025 V + 0.00004 V)
= ± 0.00029 V
Four counts on the 2 V range equals 0.00004 V.
Therefore, the accuracy range for the 1.00000 V reading is 0.99971 V to 1.00029 V.
Appendix A
Accuracy calculations

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
Input Impedance>200 TΩ
Electrometer TypeElectrometer/High Resistance Meter
Current Measurement Range20 fA to 20 mA
DisplayVacuum Fluorescent Display
Communication InterfacesIEEE-488 (GPIB), RS-232
Power Supply100 V / 120 V / 220 V / 240 V; 50 Hz / 60 Hz
Resistance Measurement RangeUp to 200 TΩ

Related product manuals