EasyManua.ls Logo

Keithley 6517B

Keithley 6517B
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 6517B Electrometer Reference Manual Section 3: Connections
6517B-901-01 Rev. C / August 2015 3-3
Connecting PREAMP OUTPUT, COMMON, or 2V ANALOG OUTPUT to earth while floating the
input may damage the instrument.
Figure 11: Maximum input levels
Input protection
The Model 6517B incorporates protection circuitry against nominal overload conditions. However, a
high voltage (>250 V) and resultant current surge could damage the input circuitry. A typical test
circuit to measure the leakage current of a capacitor is shown below. When Switch S is closed, an
initial surge of charging current flows and the high voltage is seen across the input of the Model
6517B.
Figure 12: Capacitor test circuit without protection

Table of Contents

Other manuals for Keithley 6517B

Related product manuals