Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-42 6517B-901-01 Rev. C / August 2015
CLEar[:IMMediate]
This command clears the LIMIT 1 and LIMIT 2 test failures.
Usage
:CALCulate3:LIMit[1]:CLEar[:IMMediate]
:CALCulate3:LIMit2:CLEar[:IMMediate]
Details
These action commands are used to clear the fail indication of LIMIT 1 and LIMIT 2 tests. Note that a
failure is also cleared when the limit test is disabled (:STATe OFF).
Also see
None
:AUTO <b>
This command controls auto-clear for LIMIT 1 and LIMIT 2.
Usage
:CALCulate3:LIMit[1]:CLEar:AUTO <b>
:CALCulate3:LIMit2:CLEar:AUTO <b>
:AUTO?
<b>
1 or ON: Enable auto-clear for limit failure
0 or OFF: Disable auto-clear for limit failure:I
Details
With auto-clear enabled, the fail indication of a limit test clears when instrument operation enters the
idle state. With auto-clear disabled, the fail indication remains until it is cleared by the
:CLEar[:IMMediate] command.
Also see
None