EasyManuals Logo

Keithley 6517B Reference Manual

Keithley 6517B
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #411 background imageLoading...
Page #411 background image
Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-160 6517B-901-01 Rev. C / August 2015
:STOP <NRf>
These commands specify the stop test voltages and query the stop voltage for the specific test.
Usage
:TSEQuence:DLEakage:STOP <NRf>
:TSEQuence:STSWeep:STOP <NRf>
:STOP?
<NRf>
-100.0 to 100.0 Volts; 100 V range
-1000 to 1000 Volts; 1000 V range
Details
This configuration command is used for the following tests:
Diode leakage test
Staircase sweep test
This command is used to specify the stop voltage for the specified test.
:TSEQuence:DLEakage:STOP <NRf> stops voltage; diode leakage test.
:TSEQuence:STSWeep:STOP <NRf> stops voltage; staircase sweep test.
:STOP? queries the stop voltage.
Also see
None
:STEP <NRf>
This command specifies the step voltage for a diode leakage or staircase sweep test.
Usage
:TSEQuence:DLEakage:STEP <NRf>
:TSEQuence:STSWeep:STEP <NRf>
:STEP?
<NRf>
-100.0 to 100.0 Volts; 100V range
-1000 to 1000 Volts; 1000V range
Details
This configuration command is used for the following tests:
Diode leakage test
Staircase sweep test
:TSEQuence:DLEakage:STEP specifies the step voltage for a diode leakage test.
:TSEQuence:STSWeep:STEP specifies the step voltage for a staircase sweep test.
:STEP? queries the step voltage
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
Input Impedance>200 TΩ
Electrometer TypeElectrometer/High Resistance Meter
Current Measurement Range20 fA to 20 mA
DisplayVacuum Fluorescent Display
Communication InterfacesIEEE-488 (GPIB), RS-232
Power Supply100 V / 120 V / 220 V / 240 V; 50 Hz / 60 Hz
Resistance Measurement RangeUp to 200 TΩ

Related product manuals